{ "id": "flsol_taskE_auto_optimize_scan", "snapshot": "flsol", "instruction": "一步一步地执行,别他妈的给老子跳步 完全按照老子给的指令执行5.点击'PMT_Voltage'(758,369)设置为700V;6.设置'EX_Slit(下拉框)(771,321)'为2.5nm,'点击下拉框EM_Slit(764,344)'出现下拉选项,选择5nm(748,386);7.点击'Report(638,290)'确认'Data_start(718,455)'和'Data_end(717,480)与instrument的EM_Start_WL和EM_End_WL'的值一致;8.点击'确定(641,717)';9.点击'Measure(182,162)'测量并观察谱图;10.若信号截断则返回'Method(36,166)'降低'PMT_Voltage(774,369)'或缩小狭缝(771,321)后重测;11.若信号过弱则返回'Method(36,166)'升高'PMT_Voltage(774,369)'或增大狭缝(764,344)后重测;12.重复调优至峰值在30%-90%且峰形完整;13.点击'输出结果(108,948)'并最大化。", "source": "custom", "config": [], "trajectory": "trajectories/", "related_apps": [ "flsol" ], "evaluator": { "postconfig": [ { "type": "sleep", "parameters": { "seconds": 5 } } ], "func": "vllm_eval", "expected": { "description": "FL Solutions 主界面中图表区域应显示一条完整的荧光发射光谱曲线:峰形平滑、顶部无截断(曲线最高点不贴近纵轴上限)、基线平稳、信噪比良好。界面中的仪器参数区域应可见激发波长 350 nm、发射扫描范围 380-700 nm,以及经过迭代调整后的最终 PMT 电压和狭缝宽度参数。" } }, "proxy": false, "fixed_ip": true, "possibility_of_env_change": "medium", "metadata": { "input_files": [], "steps": "{\"application\": \"FL Solutions - F-4600\", \"main_window\": {\"toolbar\": [{\"name\": \"Method\", \"center\": [36, 166]}, {\"name\": \"Measure\", \"center\": [182, 162]}]}, \"sub_windows\": {\"Analysis_Method\": {\"tabs\": [{\"name\": \"Instrument\", \"center\": [456, 293]}], \"footer_buttons\": [{\"name\": \"确定\", \"center\": [641, 717]}, {\"name\": \"取消\", \"center\": [719, 717]}], \"Instrument\": [{\"name\": \"EX_WL\", \"center\": [497, 472]}, {\"name\": \"EM_Start_WL\", \"center\": [493, 500]}, {\"name\": \"EM_End_WL\", \"center\": [495, 524]}, {\"name\": \"EX_Slit\", \"center\": [771, 321]}, {\"name\": \"EM_Slit\", \"center\": [764, 344]}, {\"name\": \"PMT_Voltage\", \"center\": [774, 369]}]}}}", "steps_original": "1. 打开 FL Solutions,新建波长扫描方法。\n2. 设置激发波长 350 nm,发射范围 380-700 nm,初始 PMT 700V,狭缝 2.5 nm。\n3. 执行测量,观察谱图。\n4. 若峰截断则降低 PMT 电压和/或缩小狭缝;若信号过弱则升高 PMT 电压和/或增大狭缝。\n5. 反复迭代测量直到峰形完整显示,截图记录最终结果。", "difficulty": "hard", "highlight": "AI 能够读取谱图质量并进行闭环迭代调参,体现真正的仪器操控智能,而非机械执行固定步骤。", "ui_coordinates": { "application": "FL Solutions - F-4600", "main_window": { "name": "F-4600 FL Spectrophotometer on USB", "toolbar": [ { "name": "Method", "type": "Button_Icon", "center": [ 36, 166 ], "description": "打开分析方法设置窗口" }, { "name": "Measure", "type": "Button_Icon", "center": [ 182, 162 ], "description": "执行测量" } ], "monitor_panel": [ { "name": "Fluorescence_Value", "type": "Display", "center": [ 959, 240 ] }, { "name": "EX_WL_Display", "type": "Text", "center": [ 912, 262 ] }, { "name": "EM_WL_Display", "type": "Text", "center": [ 0, 0 ] }, { "name": "Status_Ready_Label", "type": "Status_Indicator", "center": [ 0, 0 ], "color_hint": "Green" } ] }, "sub_windows": { "Analysis_Method": { "common_controls": { "tabs": [ { "name": "General", "center": [ 398, 291 ] }, { "name": "Instrument", "center": [ 456, 293 ] }, { "name": "Monitor", "center": [ 524, 295 ] }, { "name": "Processing", "center": [ 574, 293 ] }, { "name": "Report", "center": [ 638, 290 ] } ], "footer_buttons": [ { "name": "确定", "center": [ 641, 717 ] }, { "name": "取消", "center": [ 719, 717 ] }, { "name": "应用", "center": [ 788, 717 ], "status": "disabled" }, { "name": "帮助", "center": [ 882, 717 ] } ], "close_button": { "name": "Close_Window", "center": [ 905, 263 ] } }, "tabs_content": { "General": [ { "type": "ComboBox", "name": "Measurement", "center": [ 512, 321 ], "value": "Wavelength scan" }, { "type": "Edit", "name": "Operator", "center": [ 533, 349 ], "value": "Administrator" }, { "type": "Edit", "name": "Instrument_Model", "center": [ 0, 0 ] }, { "type": "ComboBox", "name": "Sampling", "center": [ 0, 0 ], "status": "disabled" }, { "type": "Edit", "name": "Comments", "center": [ 0, 0 ] }, { "type": "CheckBox", "name": "Use_sample_table", "center": [ 0, 0 ] }, { "type": "Button", "name": "Load", "center": [ 0, 0 ] }, { "type": "Button", "name": "Save", "center": [ 0, 0 ] }, { "type": "Button", "name": "Save_As", "center": [ 0, 0 ] } ], "Instrument": [ { "type": "ComboBox", "name": "Scan_mode", "center": [ 550, 322 ] }, { "type": "ComboBox", "name": "Data_mode", "center": [ 526, 347 ] }, { "type": "Edit", "name": "EX_WL", "center": [ 497, 472 ], "unit": "nm" }, { "type": "Edit", "name": "EM_Start_WL", "center": [ 493, 500 ], "unit": "nm" }, { "type": "Edit", "name": "EM_End_WL", "center": [ 495, 524 ], "unit": "nm" }, { "type": "ComboBox", "name": "Scan_speed", "center": [ 497, 549 ], "unit": "nm/min" }, { "type": "ComboBox", "name": "EX_Slit", "center": [ 771, 321 ], "unit": "nm" }, { "type": "ComboBox", "name": "EM_Slit", "center": [ 764, 344 ], "unit": "nm" }, { "type": "Edit_Spin", "name": "PMT_Voltage", "center": [ 758, 379 ], "unit": "V" }, { "type": "CheckBox", "name": "PMT_Voltage_Limit", "center": [ 0, 0 ], "label": "PMT Voltage 0-1000V" }, { "type": "ComboBox", "name": "Response", "center": [ 0, 0 ], "unit": "s" }, { "type": "Edit_Spin", "name": "Replicates", "center": [ 0, 0 ] } ], "Monitor": [ { "type": "Edit", "name": "Y_Axis_Max", "center": [ 455, 343 ] }, { "type": "Edit", "name": "Y_Axis_Min", "center": [ 455, 375 ] }, { "type": "CheckBox", "name": "Open_processing_after_acquisition", "center": [ 396, 437 ] }, { "type": "CheckBox", "name": "Overlay", "center": [ 395, 503 ] } ], "Report": [ { "type": "Edit", "name": "Data_start_Value", "center": [ 718, 455 ], "unit": "nm" }, { "type": "Edit", "name": "Data_end_Value", "center": [ 717, 480 ], "unit": "nm" } ] } } } } } }